Reference ISQ-Mercator pressures

Most common “advanced” issue with the ISQ-Mercator is clogging of the instrument. As clogging leads to increased pressure, the clogged piece can be identified easily by disecting the instrument, if reference pressures are known.

Table of reference pressures

All pressures were recorded on a (reasonably) cleaned instrument with flow of 0.6 ml/min and 5% ACN : 95% H2O mobile phase after pressure stabilized.

Point of disconnection

Pressure (bar)

Full system

222-223

At the ISQ entrance (at the ISQ 3-way valve)

212-213

After the UV-VIS cell (excl. UV-VIS→ISQ tubing)

200-201

After the column (excl. Column→UV-VIS tubing)

196-198

Between column and pre-column

20-21

Before the pre-column (incl. Sampler→Column tubing)

3

note1: Incl. means that the part was left in the system, giving additional backpressure note2: ISQ needle itself gave only ~3 bars of backpressure with the abovementoned conditions

From the abovementioned table, specific pressure for each instrument part at the abovementioned conditions can be derived:

Part

Pressure (bar)

Pumps, autosampler, tubing to columns

3

Pre-column

17-18

Column

176-178

UV-VIS cell + Column→UV-VIS tubing

3-4

UV-VIS→ISQ tubing (redtube)

11-12

ISQ (incl. 3-way valve)

10