GC-2010 FID Gas chromatograph septum replacement

This guide of shorthand version of the official manual. For pictures and detailed description, please go to the manual page 122. [1]

To replace the septum and reset the counter:

  • Go to the [SYSTEM] -> Maint Inj (PF menu)

  • Wait for “GC is ready for maintenance” message.

  • Remove the top lid

  • Both oven and injection port should be cooled to 50 °C (or below) and inlet pressure should be 0 kPa.

  • Unscrew the septum nut.

  • Replace the septum. Do not loose the metal needle guide which is in the septum! Do not touch the new septum by hand! (use tweezers)

  • Tighten the new septum as far as you can (handtight).

  • Return a semicircle (180°).

  • Put back the top lid.

  • select [Anal.] from PF menu. GC should flow carrier gas for 5 miutes and return to operating temperature.

  • Go to [DIAG] -> Analysis Counter, go to the analysis counter, click reset.

  • Test the system sanity by running a sample.

  • If needed do a blank run(s) to remove pollution from the replacement.