LCQ Fleet mass spectrometer - Wing4 03.421 (MS, direct injection)

This instrument is intended for advanced experiments.

How to operate the machine

Go to Operation procedure section.

Notes (FAQ)

  • Stay away of pollutants, if you plan working with pollutants or high concentrations, contact Jan.[1]

How to gain access

Contact Jan.[1]

Booking system

Instrument is not in the booking system yet. It is currently in its beta-phase.

Sample preparation

Please refer to the Sample preparation (MS, LC-MS).

Operation procedure

  • Load the reference tunefile - C:\Data\_TuneFiles\calmix_reference.LTQTune.

  • Fast-forward the syringe pump till you see a drop at the ESI head.

  • Start syringe-pump (it should be tuned to 5 μl/min).

  • Turn off the ESI, scan settings: 5 microscans, 1ms injection time.

  • Wait 2-3 minutes for the signal to stabilize.

  • Check that the injection time (IT) is 1 ms or below.

  • Measure the calmix for 2 minutes and average the signal. Calmix measurement should be saved into C:\Data\_Tunemix_check\[year_month_day_start/end]

  • Take a note in the labbook.

  • Disconnect the calmix-only red tube and do not use it for your experiments.

  • Measure what you need with any tunefile you need.

  • Safe your files into your folder - C:\Data\[your name].

  • Get the instrument clean again.

  • Measure the calmix (same procedure as at the beginning).

  • Check that things are switched off at the end.

Network storage

No network storage connected.

When something goes wrong

DON’T PANIC! ;) It happens to all of us someday. Contact Jan.[1]

Questions/Misc

If something is wrong in this guide feel free to contact Jan.[1]