ISORI ion spectroscopy instrument - Wing4 (MS-UV/VIS/IR)

This is just brief operation guide, if the guide differs from Aleks recommendation, the guide is wrong.

How to operate the machine

  • pressure in the He trap must be kept below 4e-7 mbar (channel #4 TRAP on the pressure meter)

  • Q-Delay of the IR laser for OPO-B should be slowly decreased from 400 μs to 290 μs before operation.

Follow Aleks advice. In future, guideline will be made

Booking system

private booking system - the same which is for office/labspace

Switching off the instrument

  • Set He-piezo to continuous (by this you effectively turn of He pulses).

  • Set trap RF to 0.

  • Turn off trap RF generator.

  • Turn off the laser shutter.

  • Disable all pulses.

  • Switch ISORI to Q1 MS mode and set m/z to 10.

  • Switch from “Active low” mode to “None”.

  • Set accumulation to 2000 and turn accumulation on.

  • Execute “delayed_heat.bat” file at the desktop.

  • Turn off laser shutter and pulsing.

  • Set laser Q-delay to 400.

  • Turn the laser key to off.

  • Switch off TSQ multiplier.

  • Clean the ESI source and cap it.

  • Switch off the ESI source N2 inlet (big plastic knob in the inlet line above your head).

  • Check the pressures to verify that He pulses are off and that the source is capped.

  • Clean the tables around ISORI.

Network storage

Windows XP PC. No network storage connected.

Retrieving data

  • Turn off the Eclipse.

  • Wait for the Eclipse to fully close.

  • Insert your USB stick and copy your data.

  • Safely remove your USB stick.

  • Reboot the PC.

  • Turn on the Eclipse.

  • Check that Eclipse started normally.

When something goes wrong

Contact Aleks! If Aleks is not around, check with Guilherme or Jaya

Questions/Misc

If something is wrong in this guide feel free to contact me.